Microscope set-up and resolution

Improve a picture of a plant specimen by setting up the microscope the way the lessons teach. See what the numbers (resolution, depth of field, useful magnification) do to what you can actually see.

Loading the lab…

About this model

Resolution is set by the numerical aperture (NA), not by magnification. With the condenser in use, the smallest gap that can be separated is d = 1.22 λ ÷ (NA of the objective + NA of the condenser), which is 0.61 λ ÷ NA when they match; at 550 nm and NA 0.95 that is 0.35 µm (the Abbe form λ ÷ 2NA gives 0.29 µm). Closing the condenser aperture raises contrast and costs resolution; the usual compromise is 70 to 80% of the objective NA. Depth of field is λ n ÷ NA² + n e ÷ (M NA), so it shrinks fast with NA. Useful magnification is about 500 to 1000 times the NA; more is empty magnification. In epifluorescence, brightness goes as NA⁴ ÷ M².

Contrast is C = (background − specimen) ÷ background. Stained, absorbing structures show in brightfield (a transmission of 0.40 gives C = 0.60); unstained cells are phase objects with almost no brightfield contrast (about 4% for a phase shift of 0.2 radians) and need phase contrast (a ring transmitting 25%, so C is about 0.64), where bright halos around structures are artefacts. In fluorescence the filter cube must excite the dye and pass its emission; a band that also passes chlorophyll adds bleed-through, and long or strong excitation bleaches the dye.

The pictures are drawn by a simple model: a Gaussian blur whose width reproduces the Rayleigh dip of about 73.5% between two points, plus defocus, veiling glare, uneven illumination and noise. Assumptions the lessons do not give, chosen for teaching: a viewer needs at least 5% contrast (20% is good) or a signal-to-noise ratio of 5 (10 is good); veiling glare grows as 0.8 times the square of the aperture; brightfield contrast of unstained objects rises as the aperture closes; an incomplete Köhler set-up cuts contrast by 30% (condenser not focused) and 10% (not centred); the spectra of the dyes are Gaussian bands around their usual peaks (GFP 488 and 509 nm, chlorophyll emission 685 nm, Calcofluor white 355 and 435 nm, lignin 380 and 470 nm); bleaching rates are teaching constants. The model assumes the correct coverslip (0.17 mm) and an infinity-corrected, parfocal objective set.

Calibrated against: PMI-201 lessons 1.3, 1.4, 2.1, 2.2, 3.1, 3.2 and 3.4 (the worked examples, including 1.34, 0.52 and 0.26 µm; 516, 583 and 605 nm; 11.4, 1.55 and 0.57 µm; contrast 0.60; I ÷ I0 = 0.36; brightness ratios 4.6, 4 and 5, are reproduced in the unit tests); Abbe 1873; Rayleigh 1879 (resolution limits); Köhler 1893 (illumination); Zernike 1942 (phase contrast).

This is a teaching model. It shows how the variables behave and what to expect; your own instruments, reagents and material will differ, so always follow your lab's validated procedure.

Related lessons: Light, NA and resolution (PMI-201 lesson 1.3),Köhler illumination (PMI-201 lesson 2.2),Contrast and fluorescence (PMI-201 module 3)